Technical Data  
   
Pressure Range:  0,1-10 bar
Pressure Fall:  5 Pa
Reaction Time: 20 ms
Measuring Range 1: 0-500 Nmm3/s
  0-30 Nml/min
Measuring Range 2: 0-5 Ncm3/s
  0-300 Nml/min
Accuracy: 2% of max. scale value
Display: Flow value in mm3/s or ml/min
Flow connection:  Hose with quick coupling for connecting to reference leaks in Nolek instruments.
Battery:     2 x AA Batteries
Casing:  Aluminium profile
170x80x35mm, 390g
Additional Features:    Automatic shutoff
  Memory of flow value
  PC Communication available
  Rubber protection to minimise damage and electrical shock

 Applications

 

C9 is used for accurate calibrating and setting of air flows from adjustable reference leaks in leak testing or flow control instruments.

 

C9 is pressure independent and measures the flow for over pressure as well as vacuum conditions.

 

C9 is a user friendly calibration unit. When operated within the pressure and flow ranges stated under Technical Data, the resultant airflow is shown immediately in mm3/s or ml/min in the display.

 Function

 

The unit has a mass flow meter with display and hose with quick coupling for connecting to the instrument to be tested.

 

For over pressure conditions, the hose is connected to the right connection of C9, and for vacuum conditions, to the left connection and to the instrument's reference leak.

 

When the flow from the instrument has stabilised, the value is indicated on the display.

 

TQC Ltd, Hooton Street, Carlton Road,
Nottingham, NG3 2NJ, United Kingdom
Tel: +44 (0)115 9503561