leak test facility
for small
medical device
The facility is designed to leak test a
sealed medical device. The test is to a limit which is intended to ensure that no inward
leakage will occur at pressures up to 0.5 bar differential. An air leakage limit of the
order of 20 mm3/sec is considered to meet this requirement and the actual equipment is set
to operate below this value.
The
medical devices are sealed with an enclosed internal volume, therefore the test method is
to enclose the capsule in a pressurised chamber and measure any pressure decay due to
leakage in the capsule. this is done using the standard air decay sequence of a nolek
TS600 leak test instrument. However, in order to check for gross leaks (which would
equalise pressure internally and therefore invalidate the test) a dosing system is added
to the basic instrument (the new designation being TS624)
The test fixture is a simple bench mounting
unit with a slide in drawer containing the lower half of the test chamber. The upper half
is clamped by a pneumatic cylinder which can only operate when the drawer is fully closed.
Once the drawer is closed, an interlock valve is operated which allows a pneumatic signal
from the instruments auto connector output to operate the cylinder via a pilot operated
valve.
TQC welcomes the opportunity to design and
manufacture complete medical leak test systems to client specifications, whether you have
a simple bench top application or a more complex automated leak test requirement. Contact
our engineers to advise you on your automation project.
|
TQC Ltd, Hooton Street, Carlton Road,
Nottingham, NG3 2NJ, United Kingdom
Tel: +44 (0)115 9503561
Fax: +44 (0)115 9484642
E-mail: sales@tqc.co.uk
|
|